Contact an Account ManagerSeparation of β-lactam antibiotics on NUCLEODUR® C18 Gravity
Summary
Type: HPLC
Application #: 124942
Catalog#: 760103.40 - MN HPLC Column, Nucleodur C18 Gravity, 110A, 5um, 4x150mm
Detection:
Conditions: Eluent A: acetonitrile
Eluent B: 20 mmol/L KH2PO4, pH 3.5
Gradient: 10 % A for 3 min, 10–50 % A in 9 min, 50 % A for 3 min
Flow rate: 1.0 mL/min
Temperature: 25 °C
Injection volume: 1.0 µL
Authors: HPLC department
Source: Macherey-Nagel, Germany 2011
©2020 Copyright, All Rights Reserved, Sorbent Technologies, Inc.